Non-contact Surface Roughness/Profile Measuring Instruments
Non-contact surface roughness/profile measuring Instruments
A non-contact measuring instrument uses light in place of the stylus of used in a contact-type measuring instrument. These instruments come in multiple types, such as confocal and white light interference, and vary depending on the principle used. There are also a variety of contact-type detectors that have been changed into non-contact instruments by replacing the probe with optical sensors and microscopes. We will use KEYENCE's 3D laser scanning microscope, the VK-X Series, as an example to explain the principles of confocals.
A 3D laser scanning microscope uses the confocal principle, and a laser as the light source, to measure the asperity of the target's surface.
The system is configured as shown in the figure on the right. Set the sample on the XY stage of the measurement unit and perform the 3D scan.
With KEYENCE's VK-X Series, an X-Y scanner is embedded in the measurement unit. The laser light source scans across the surface of the target in the X and Y directions and acquires the surface data.
The scanning principles are explained below.
3D laser scanning microscope measurement principles
- 1. The laser beam emitted from the laser light source scans the target surface.
- 2. The laser light is reflected from the target surface, passes through the half mirror, and enters the light-receiving element. At this point, the laser intensity of the received reflection, as well as the height position of the lens, are recorded by the microscope. The laser microscope acquires 1024 data points in the X direction and 768 data points in the Y direction, and records the intensity and lens height for each point (1024 x 768 = 786432 points).
- 3. When the scan of one surface finishes, the objective lens moves in the Z direction by the specified pitch.
- 4. The same surface scan is performed again for the surface that the objective lens has moved to, and the laser’s reflected light intensity is checked over 1024 × 768 points. The reflected light intensity of each pixel is compared with the reflected light intensity recorded in memory as “v”. If the new reflected light intensity is higher, the reflected light intensity data and lens height position data are overwritten.
- 5. The operations of steps (2) to (4) are repeated for the specified Z distance.
- 6. Finally, for each of the 1024 × 768 pixels, the reflected light intensity and lens height position are recorded in memory at the time when the strongest laser light reflection was received.
- 7. For optical microscopes, the WD (Working Distance: the distance from the objective lens to the target) when the objective lens is in focus is constant. If it is assumed that the image is in focus when the reflected light intensity is at maximum, it is possible to obtain 3D data in the observation area (1024 × 768 pixels) of the microscope by stitching together the lens height positions from the different times that the image was in focus, that is, when the reflected light intensity was at its maximum.
3D laser scanning microscope accuracy
The ability to accurately read the peak value of the reflected light intensity has a large effect on the measurement accuracy of laser confocal measurement systems.
There are many ways to construct a confocal optical system. The pinhole confocal method used in KEYENCE's 3D laser scanning microscope is explained below:
With the pinhole confocal method, a pinhole is placed in front of the light-receiving element. The pinhole has a diameter of just tens of micrometers and has the role of blocking reflected light when the image is not in focus.
When the image is in focus (see figure below), the reflected light is received by the light-receiving element in both the normal optical system and the laser confocal optical system. When the image is not in focus, the reflected light (out-of-focus light) enters the light-receiving element for normal optical systems, but is blocked by the pinhole when using the pinhole confocal method. In other words, the structure is such that the reflected light only enters the light-receiving element when the image is in focus.
The pinhole's effect on the received light is illustrated in the figure on the right. With the laser confocal, the reflected light intesntiy peaks at the focal point. On the other hand, the normal optical system results in a gently sloping curve.
The lack of a peak at the focal point makes it difficult to detect when the target is in focus.
Laser XY-direction resolution
For non-contact systems, the light beam spot corresponds to the stylus of contact-systems. Non-contact systems do not directly touch the target, therefore they do not have the disadvantages of stylus wear and the risk of scratching a sample. The size of the beam spot diameter is important in order to accurately measure the profile of a sample. Generally, the smaller the beam spot diameter, the smaller the features that can be measured.
Laser microscopes use lasers for their light sources, which makes it possible to create an extremely small beam spot.
When using a 150x (N.A. = 0.95) objective lens, the VK-X Series, which uses a 404 nm violet laser for its laser light source, achieves a resolution of 0.13 μm 0.005 Mil for the planar spatial resolution. A laser microscope can measure asperity with a very small width, which cannot be measured with contact-type systems.
by David Beamish, DeFelsko Corporation
Updated: 09/20/2021
Abstract: Coating performance is related to the profile height on a steel surface. Three types of devices are available to take measurements of this surface profile: replica tape, depth micrometers fitted with pointed probes, and stylus roughness testers. This paper presents results from a recent analysis of measurements taken by the three device types on steel blasted with an assortment of blast media and proposes a new method of depth micrometer measurement called average of the maximum peaks.
Introduction to Surface Profile Measurement
Steel surfaces are frequently cleaned by abrasive impact prior to the application of protective coatings. This process removes previous coatings and roughens the surface to improve coating adhesion. The resultant surface profile, or anchor pattern, is comprised of a complex pattern of peaks and valleys which must be accurately assessed to ensure compliance with job or contract specifications.
Protective coating professionals have several testing methods available to them for determining the amount of surface profile. Little information has been available to help them select an instrument or compare results from different methods.
Measurement Methods—How is Surface Profile Measured?
A steel surface after blast-cleaning consists of random irregularities with peaks and valleys that are not easily characterized. Instruments that can measure this profile with a high degree of precision, such as scanning electron microscopes, are suitable only for laboratory use. Field methods are desirable. Surface profile ranges are frequently specified and the recommended surface profile is different for various types of coatings.
The determination of surface profile depends on its definition. ISO1 8503-12 defines it as the height of the major peaks relative to the major valleys. ASTM3 D71274 describes it as the positive and negative vertical deviations measured from a mean line, approximately the center of the profile being evaluated. ASTM D4417-115 defines surface profile as, "the height of the major peaks relative to the major valleys". It describes 3 different measurement methods:
- Method A—profile comparators
- Method B—depth micrometers
- Method C—replica tape
The industry does not have profile standards with values traceable to a National Metrology Institute. If they did, instruments could be verified against those standards, accuracy statements could be published, and users would have a means of correlating their results. Standards could determine the relationship of values obtained from replica tape to those obtained from depth micrometers, and so on.
Not having physical standards, the industry has chosen a referee method. NACE6, ASTM, and ISO describe surface profile height as distance measured from the top of the highest peak to the bottom of the lowest valley in the field of view of an optical microscope. A microscope is focused on the highest peak within the field of view. The distance travelled by the lens in order to focus on the lowest valley within the same field of view is a single measurement of profile height. The arithmetic mean of 20 such measurements results in the mean maximum peak-to-valley height. In other words, the average of the maximum peaks.
Fig.2 Computer generated image of a blast cleaned steel surface (left). A blasted surface (right)The microscope method is impractical in the field, so major organizations support a number of alternative methods that are both practical and routinely used by inspectors.
ISO manufactures surface profile comparators for steel blast-cleaned by shot or grit abrasives7 that are based on the focusing microscope method. Using visual or tactile means, the user compares the steel surface with the profile of each segment of the comparator to apply an appropriate grading of “fine”, “medium” or “coarse”. Annex B of ISO 8503-5 shows there is good correlation between these comparators and measurement by replica tape and stylus methods. There is no ISO method for depth micrometers nor should depth micrometers be used to measure on profile comparators due to the lack of flatness of the comparators.
NACE RP0287 (updated in 2016 to SP0287-2016-SG) also shows8 replica tape and focusing microscope measurements agree within their confidence limits (two standard deviations) in 11 of 14 cases.
Fig.3 Replica TapeHow Replica Tape Readers Measure Surface Profile
Replica tape is simple, relatively inexpensive, and shows good correlation to focusing microscope results. It is not surprising then that it has arguably become the most popular field method for measuring surface profile.
Replica tape consists of a layer of compressible foam affixed to an incompressible polyester substrate of highly uniform thickness (2 mils +0.2 mils9). When pressed against a roughened steel surface, the foam collapses and forms an impression of the surface. Placing the compressed tape between the anvils of a micrometric thickness gage and subtracting the contribution of the incompressible substrate, 2 mils, gives a measure of the surface profile.
Automatically subtract the 50.8 μm (2 mils) incompressible film from all readings with the PosiTector RTR H Replica Tape Reader.
Additional reading:What is Voltage Testing?
Types of Drone Test Thrust Stands
Key Questions to Ask When Ordering Igu-16 Single-Component Intelligent Seismic Sensor: A Comprehensive Guide
How Does Drone Photography Work?
Unlocking the Power of Seismic Learning Recorder
According to ISO 8503-5 “This method measures an ‘average maximum peak-to-valley profile’ because the anvils of the micrometer gauge flatten the replica profile slightly so that the reading equates to an average maximum value, though this is not the same as a mathematical average.” So again, we have a method that essentially measures the average of the maximum peaks.
In recent years, two other methods of profile measurement have gained popularity: the stylus roughness tester (ASTM D7127) and the depth micrometer (ASTM D4417 Method B). Electronic versions of these instruments have the advantage of reduced operator influence and digital collection and analysis of measurement data.
For more information on digital surface profile instruments, please see the PosiTector SPG Digital Surface Profile Gauge or the PosiTector RTR H Digital Replica Tape Reader.
How Stylus Roughness Instruments Measure Surface Profile
A portable stylus surface roughness measuring device operates by drawing a stylus at constant speed across the surface. The instrument records the up and down distances the stylus travels as it traverses across the surface. It measures Rt in compliance with ISO 428710 wherein Rt is the vertical distance between the highest peak and lowest valley within any given evaluation length of 0.5 inches. Five of these traces are made and Rt values averaged to again obtain the average of the maximum peaks.
Fig.4 Stylus Roughness Instruments (Instruments shown are similar to the ones used in this study)ASTM Committee D01.46 Round Robin Assessment of Replica Tape Readers and Stylus Roughness Instruments
ASTM committee D01.46 completed an 11 laboratory round robin assessment of precision and bias for this method by having participants measure five blasted steel test panels with replica tape and three stylus instruments. They selected stylus instruments that had adequate vertical range to be useful for measurement of the comparatively rough surfaces of interest to the coatings and linings industry. Even so, the profile on some of the panels exceeded the measuring limits of some of the selected instruments.
Preliminary findings confirmed a close relationship between replica tape and stylus roughness methods just as ISO concluded. When results are published, industry professionals will have access to reliable correlation data.
That leaves only the depth micrometer method without a comparison study. To provide correlation between all three device types, this paper proposes that depth micrometer measurements are analyzed using a method that produces results similar to tape and stylus results and is consistent with their measurement objectives, a method called "average of the maximum peaks".
To obtain this value, the profile is measured at a sufficient number of locations to characterize the surface, typically five. At each location, ten readings are taken and the highest reading is recorded. The average (mean) for all the locations is reported as the profile of the surface.
Impetus for this study came from preliminary testing on ASTM panels with a single depth micrometer instrument. As shown in figure 5, when the average of the maximum peaks method of analysis was used, depth micrometers results aligned closely with tape and stylus results.
Fig.5 Preliminary Results on 5 ASTM PanelsHow Depth Micrometers Measure Surface Profile and How They Compare to Replica Tape Readers and Stylus Roughness Instruments
A depth micrometer instrument has a flat base which rests on the surface and a spring-loaded probe which drops into the valleys of the surface profile. The flat base rests on the highest peaks and each measurement is therefore the distance between the highest local peaks and the particular valley into which the tip has projected.
Fig.6 Depth Micrometers (Instruments shown are similar to the ones used in this study)
Currently, ASTM D4417 requires the user to average all depth micrometer measurements regardless of how low some readings might be. Not surprisingly, final calculated results are usually less than what is obtained by tape and stylus methods. This study confirmed that assumption (Fig.12). Occasionally one of the instruments would register values at or above tape results, but that was the exception.
After the ASTM 5-panel study referenced above, the depth micrometer method was the only method without a comparison study. To provide correlation between all three device types, this paper proposes that depth micrometer measurements are analyzed using a method that produces results similar to tape and stylus results and is consistent with their measurement objectives, a method called "average of the maximum peaks".
To obtain this value, the profile is measured at a sufficient number of locations to characterize the surface, typically five. At each location, ten readings are taken and the highest reading is recorded. The average (mean) for all the locations is reported as the profile of the surface.
Impetus for this study came from preliminary testing on ASTM panels with a single depth micrometer instrument. As shown in figure 5, when the average of the maximum peaks method of analysis was used, depth micrometers results aligned closely with tape and stylus results.
Summary of Test (to Compare Depth Micrometers to Replica Tape Readers and Stylus Roughness Instruments)
To confirm these results, twenty panels blasted with common media types were obtained from KTA Labs11 and five common depth micrometers were acquired. Five individuals took 50 measurements on each panel with each instrument in a controlled office environment for a total of 5,000 readings.
A minimum of 3 replica tape measurements were taken on each panel and averaged. When results fell in the outer regions of the tape’s range, additional measurements were obtained with the next level of tape and averaged as per the manufacturer’s instructions.
See "Replica Tape - A Source of New Surface Profile Information" for more information about Replica Tape Measurement.
Stylus roughness measurements were obtained from three common field instruments for comparison. Finally, base metal readings (BMR) from each panel were obtained from Type 1 and Type 2 magnetic coating thickness gages.
Fig.7 Panel Measurement Locations for each MethodSurface Profile's Effect on DFT (Coating Thickness) Instruments
DFT probes measure the distance from their probe tip to the magnetic plane in the steel. On smooth steel the magnetic plane is coincident to the surface of the steel. On rough steel the magnetic plane lies somewhere between the highest peak and the lowest valley in the profile, a location that may differ by instrument type. Therefore roughness generally causes DFT instruments to read high, or a positive value.
SSPC-PA 2 and other standards require a correction factor be applied in order to compensate for this roughness effect. Commonly, a plastic shim is placed over the bare profile and measured with the DFT gage. The gage is adjusted so that the result matches the thickness of the shim. The shim simulates the paint build-up over the peaks and the adjustment ensures paint thickness measurements are taken from the average level of the peaks of the profile, rather than from the magnetic plane.
To quantify the effect of profile on DFT gages, measurements were taken on all panels by Type 1 (mechanical pull-off) and Type 2 (electronic) instruments after first being zero-checked on smooth, flat steel. The average result of five measurements was recorded for each panel.
The Type 1 instrument was least affected by profile and measured a maximum of 0.3 mils on the roughest surface. The Type 2 instrument measured between a low of 0 on the glass bead blasted surface and a high of 1.2 mils on the S390 shot blasted surface. Overall, the DFT instrument gave thickness results that ranged between 1 and 26% of surface profile heights as measured by replica tape, with an average of 13% across all panels.
Fig.8 DFT Gage Results Compared to Replica Tape ResultsGeneral Observations of Surface Profile Measurement
Some surface roughness exceeds the measuring capabilities of tape and stylus methods. Good practice suggests that commercial grades of tape permit measurement of average peak-to-valley profiles of between 0.5 to 5.0 mils. All depth micrometers used in the study had extended ranges suitable for measuring blasted steel surfaces and did not "max" out on any of the panels.
View the PosiTector SPG Surface Profile Gauge ordering guide for measuring ranges.
Several panels had areas where all instrument types produced high profile values. These variances might have been due to the inconsistent nature of blasting by hand. It can be assumed larger surfaces would have similar irregularities.
It was not possible to test with each device in the exact same location on each panel (Fig.7). Replica tape examined a relatively large area thus requiring fewer measurements to adequately characterize the surface. Stylus and depth micrometer methods have fine-tipped probes which sample a much smaller surface area and therefore required more measurements to adequately characterize a surface. ISO, ASTM, NACE and SSPC guides take this into account.
All methods required initial setup and accuracy verification before testing began.
Refer to the PosiTector SPG and PosiTector RTR H instruction manuals to learn about setup and accuracy verification.
- The replica tape method required micrometer accuracy be checked against a known thickness such as a plastic shim and its dial set back 2 mils to account for the non-compressible plastic layer. Minor adjustments had to be made during the test to compensate for micrometer drift.
- Stylus roughness testers required the most setup. The proper evaluation length was input, reporting parameters such as Rpc (peak count) and Rt (max peak-to-valley height in an evaluation length) were established, and the instrument had to be positioned with care on the blasted steel surface.
- Depth micrometers were checked at zero on a glass plate and on a shim of known thickness before and after each set of 50 measurements. No instrument drifted from zero throughout the test.
Circles were observed on some panels after testing with replica tape. It is believed they were the result of microscopic particles getting impressed into the foam and being carried away when the foam was peeled off. Scratches were observed on some panels after testing with the stylus instruments. It is believed the steel surface was slightly modified as the diamond-tipped stylus was dragged over the peaks (Fig.9).
Fig.9 A 400x Magnified Photo of Garnet-blasted Steel with a ScratchIt becomes clear during testing that individual surface profile measurement results are less repeatable and have greater variation than users have come to expect from other forms of industry measurement such as dry film thickness (DFT), temperature or gloss testing. While two DFT measurements might be expected to be very close, two surface profile measurements can differ considerably. Such is the nature of a blasted surface.
For example, on a panel blasted with a mixture of coarse and fine staurolite sands, replica tape measurements ranged between 1.8 and 2.9 mils, stylus instruments between 1.8 and 2.8 mils, and depth micrometers between 0 and 5.6 mils. Yet all three methods gave final “average of the maximum peaks” results of approximately 2.5 mils.
Just as often, however, the three methods yielded results that were not as close. Tape and stylus results sometimes varied by as much as 30%. On 2 panels blasted with S280 shot and #100 mesh aluminum oxide, replica tape read 2.7 mils on both while the stylus method averaged a lower 2.2 mils on both. Conversely on BX-40 silica sand, replica tape read 1.5 mils while the stylus method averaged a higher 1.9 mils. The average values obtained from three stylus instruments were higher than replica tape values on all 4 sand blasted panels and lower on all oxide and shot blasted panels. See figure12 for a summary of replica tape vs. stylus results.
Observations of Depth Micrometer Measurement
The following points were observed when performing surface profile measurements with the depth micrometers:
- Loose Surface Contaminant: Several panels generated high outlier measurements that were not used in the final analysis. Participants reported the instruments “rocked” on the surface. This alerted them to the issue of surface contaminants and so they avoided those areas.
- Reading Variations: There was less measurement variation on sand blasted panels compared to panels blasted with glass beads. Of 250 measurements taken with one instrument on a 4"x6"x1/8" panel blasted with Garnet, results ranged from 0.2 to 1.9 mils. When only the highest readings were averaged, the 1.2 mils result was close to tape and stylus results.
Low readings near zero were occasionally recorded. They were likely caused when a large peak pushed the probe tip up near the plane of the instrument’s foot. Averaging only the maximum values prevents these low readings from influencing the final outcome.
The highest reading in the above example of 1.9 mils is also of interest. It would seem to indicate a single, deep valley into which the probe tip descended, a large peak in the profile that elevated the depth micrometer’s foot, or surface waviness. Either way, it was only one result out of many that were averaged to obtain a meaningful profile measurement. - Number of Measurements for Analysis: When only 3 readings were taken in each location on the panels, results did not correlate closely with tape results, suggesting an insufficient number of readings. When 5 readings per location were used, final results were closer to tape results. Increasing the number of readings to 10 per location (per ASTM) removed the apparent randomness in the results and provided best correlation with tape and stylus methods. More measurements did little to improve results.
Reducing the number of locations from 5 to 3 made little difference to the overall results. This suggests a minimum of 10 readings in each of 3 locations sufficiently characterizes a blasted profile surface. - Difference in Results among Depth Micrometers: The depth micrometers used in this study had probe tips that were machined at 30° and 60° included angles. Their spring pressures were between 70 and 125g force. Instruments with 30° probes often produced lower results than instruments with 60° probes. Instruments with weak probe forces generally produced lower results than instruments with strong probe forces. This suggests probe tip angle and probe tip force affect measurement results (Fig.10).
High resolution photos of the probe tips were examined. All tips properly measured 30 or 60° as advertised but their tip radii varied considerably. Some were properly rounded. Others showed flattened or chiseled ends (Fig.11).
- Analysis Methods: When 50 readings from each depth micrometer are averaged according to ASTM D4417, the resultant profile height measurements were almost always lower than both tape and stylus. When only the maximum values from each location were averaged the results better correlated with both tape and stylus (Fig.12).
Conclusions and Deductions
Results from this study confirm a close relationship between tape and stylus measurements as first shown by ASTM round robin testing. The results also revealed interesting information about the third measurement device type, surface profile depth micrometers, which achieved results comparable to tape and stylus when “average of the maximum peaks” analysis approach was used (Fig.12).
The surface of blasted steel at any point is a random variation, so a number of readings must be taken. The assessment objective is to make maximum peak to valley determinations. Individual measurements of the surface of an abrasive blast-cleaned metal surface vary significantly from area to area over a given surface. How these measurements are combined depends on the parameter required for the job which could be the average peak-to-valley height, its maximum, or even something else. By employing the “average of the maximum peaks” analysis approach, a depth micrometer gives reliable surface profile measurements that correlate closely to replica tape and stylus roughness tester results.
PosiTector SPG Advanced models feature a SmartBatch™ mode to comply with various standards and test methods. By default, SmartBatch™ generates results close to those obtained with replica tape and drag stylus methods by automatically averaging the maximum profile depth for all spots within the test area and displaying "the average of the maximum peaks".
Citations
- International Organization for Standardization (ISO), 1 rue de Varembé, Case postale 56, CH-1211, Geneva 20, Switzerland
- Preparation of steel substrates before application of paints and related products — Surface roughness characteristics of blast cleaned steel substrates — Part 1: Specifications and definitions for ISO surface profile comparators for the assessment of abrasive blast-cleaned surfaces
- ASTM International, 100 Barr Harbor Drive, West Conshohocken, PA 19428
- ASTM D7127 “Standard Test Method for Measurement of Surface Roughness of Abrasive Blast Cleaned Metal Surfaces Using an Electronic Portable Stylus Instrument” (West Conshohocken, PA: ASTM)
- ASTM D4417 “Standard Test Methods for Field Measurement of Surface Profile of Blast Cleaned Steel” (West Conshohocken, PA: ASTM)
- From NACE Standard RP0287-2002, “Field Measurement of Surface Profile of Abrasive Blast-Cleaned Steel Surfaces Using a Replica Tape”. (Houston, TX: NACE, 2002)
- ISO 8503-2 Preparation of steel substrates before application of paints and related products — Surface roughness characteristics of blast-cleaned steel substrates — Part 2: Method for the grading of surface profile of abrasive blast-cleaned steel — Comparator procedure
- Results of NACE task group T-6G-19 round robin tests. NACE Technical Committee Report 6G176 (withdrawn). “Cleanliness and Anchor Patterns Available Through Centrifugal Blast Cleaning of New Steel” (Houston, TX: NACE International). (Available from NACE International as an historical document only.)
- This statistical summary was conducted using Imperial units. To convert to metric units, use 1 mil = 25.4 microns(μm).
- ISO 4287: 1997 Geometrical Product Specifications (GPS)—Surface Texture: Profile Method—Terms, Definitions, and Surface Parameters
- KTA-Tator, Inc. (KTA), 115 Technology Drive, Pittsburgh, PA 15275 USA.
DAVID BEAMISH (1955 – 2019), former President of DeFelsko Corporation, a New York-based manufacturer of hand-held coating test instruments sold worldwide. He had a degree in Civil Engineering and more than 25 years of experience in the design, manufacture, and marketing of these testing instruments in a variety of international industries including industrial painting, quality inspection, and manufacturing. He conducted training seminars and was an active member of various organizations including NACE, SSPC, ASTM and ISO.
Comments